71
Gases for the Avio 500 ICP
general, provide sufficient protection, but additional side shields will ensure a
further margin of safety. Safety glasses will also provide mechanical protection
for the eyes.
•
ICP-OES instruments generate high amounts of radio-frequency energy in
their RF power supply and torch boxes, which is potentially hazardous, if
allowed to escape. Safety devices and screening interlocks should not be
bypassed or disconnected.
•
The power supply of an ICP-OES is capable of generating potentially lethal
voltages. No maintenance beyond what's described in the User Hardware
Guide and Service Manual should be performed by anyone other than a
PerkinElmer Customer Support Engineer or the customer’s own PerkinElmer-
trained maintenance personnel.
•
Water lines should be located away from electrical connections. Condensation
and possible leaks may create an unsafe situation, if in proximity to electrical
connections.
The above suggestions do not supersede the safety standards outlined by OSHA or
other local state and/or country safety organizations governing safe compressed
gas cylinder handling and laboratory safety practices.
Gases for the Avio 500 ICP
Argon is used as the ICP torch gas with the Avio 500. Nitrogen is recommended
for the optical purge gas, although argon can be used as well. The quality criteria
for the argon and the nitrogen are shown in Table 2.
Table 2-2.
Table 2. Argon and Nitrogen Quality Criteria
Specification
Argon
Nitrogen
Purity
99.996%
99.999%
Oxygen
5 ppm
5 ppm
Water
4 ppm
5 ppm
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......