Peristaltic Pump
Installing the Sample Tubing
To install the sample tubing refer to the instructions below and the next two figures:
1. Turn off the peristaltic pump using the software control.
2. Take a piece of 0.76 mm (0.030 in.) pump sample tubing. New pump tubing
will work better if it is pre-stretched, and analytical precision will be improved.
Grasp the tubing, with one clip in each hand, and stretch the tubing gently a
few times.
3. Install the pump tubing on an open channel of the pump. To do this, carefully
stretch the new tubing around the pump head and place the clips on the pump
tubing into the slots, using the slots in the same position on each side of the
pump head. Carefully thread the pump tubing through the lip of the cover (see
the next figure). Be careful not to twist or kink the pump tubing during this
process.
4. Take two pieces of 0.58 mm i.d. (0.023 in.) polyethylene tubing (Part No.
09908265), one for the sample capillary and one for the nebulizer capillary.
Using a razor blade, bevel one end of each piece of capillary tubing by cutting
it at an angle. See the next figure.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......