246
Maintenance
1. Place the shear gas assembly back into position.
2. Use a T20 screwdriver to secure the two top bracket screws and use a T10
screwdriver to tighten the bottom bracket screw to secure the shear gas nozzle
3. Reinstall the torch assembly; see the section
Replacing the Torch on page 218
earlier in this chapter.
4. Align the torch and perform an axial and radial alignment; refer to the
procedure Setting the Torch Viewing Position on page 169.
Note
Before you perform an analysis at wavelengths below 190 nm, purge the
optics overnight on high purge. For other analyses, purge the system until
you obtain a steady signal.
Table 5-4.
Replacement Parts: Purge Components
Part No.
Description
09992731
Axial Purge Window
N0810346
Radial Purge Window
N0810439
Optical Kit (includes windows and O-rings)
N0810194
Axial Mirror Kit
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......