Maintenance
Installing the GemCone Nebulizer on the Scott-Type Spray Chamber
To install the GemCone nebulizer on the Scott-type spray chamber, the end cap
assembly (Part No. N0680504) and the tubing assembly (Part No. N0770336) must
be used.
1. Remove the existing nebulizer/end cap assembly (if the wrong one is installed)
by loosening the knurled nuts on the end cap. Use a gentle twisting action to
remove the assembly from the spray chamber.
2. Assemble the correct end cap (Part No. N0680504) to the spray chamber with
a twisting action.Alternately hand-tighten the knurled screws until the end cap
ring is pulled snugly against the end cap.
3. Before inserting the GemCone nebulizer into the cap, connect the argon gas
fitting to the threads on the nebulizer where shown in the next figure. First,
tighten the gas fitting by hand. Tighten the fitting another 1/8 of a turn using a
9/16-in. open-end wrench. Support the nebulizer with a back-up wrench when
tightening or loosening fittings. Do
not
overtighten the fitting as this may
cause leaking and poor ICP performance.
4. Connect the capillary tubing to the nebulizer. Be sure not to crimp or bend the
tubing as you insert it. Crimps may reduce flow to the tip of the nebulizer or
may cause air to enter the liquid stream during operation.
5. Insert the nebulizer as far as it will go through the opening in the end cap.
6. Tighten the nut on the end cap so that it is finger-tight.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......