257
GemTip Cross-Flow Nebulizer
GemTip Cross-Flow Nebulizer
The following procedures cover the maintenance of the GemTip Cross-Flow
nebulizer supplied with the instrument.
To check the performance of the nebulizer, aspirate a standard and note the
intensity reading. If the intensity is significantly lower than it should be, first check
that:
•
The pump tubing is in good condition (has no flat spots).
•
The sample capillary is in good condition (has no kinks) and is clean.
•
The plasma viewing position is set properly (using the software's
alignment function).
Also try flushing out the nebulizer by aspirating for several minutes using either:
•
a 1% acid solution, if you have been analyzing aqueous solutions, or
•
a solvent, if you have been analyzing organic solutions.
Recheck the intensity of the standard. If the intensity is still too low,
you should
check the nebulizer spray pattern. Before removing the nebulizer/end cap to check
the spray pattern, be sure to pump deionized water for several minutes to
thoroughly flush it out.
When checking the spray pattern of the nebulizer pump only use
deionized water. Do
not
use any other solution.
Avertissement:
Lors de la vérification du motif de pulvérisation de la
pompe nébuliseur utiliser uniquement de l'eau déminéralisée.
Ne pas
utiliser toute autre solution.
!
Warning
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......