167
Fitting and Connecting the Autosampler Sampling Probe
Adjusting the Sampling Probe Height
1. Install the rinsing-port rack on the autosampler.
See the procedure
Assembling Sample Racks
starting on page 61
i
n the
S10
Autosamplers User’s Guide (Part No. 09931133)
for more information.
2. Place the rinsing port in the rinse location in the sample rack.
See the procedure
Installing the Rinsing Port
on the Rack
starting on page 55
in the
S10 Autosamplers User’s Guide (Part No. 09931133)
for more
information.
3. In the ICP software, from the main menu make the following selections:
4. If the sample probe is suspended over the rinse position select the
Instrument
tab -> Load Tray
to lower the probe. If the sample probe is not at the rinse
position select the
Instrument tab -> Go To Wash
to move it there.
5. Carefully push the sampling probe down until it bottoms in the rinsing port.
Then pull the sampling probe up by about 15mm as shown in the following
figure. Use the O-ring on the probe to prevent it from slipping down. The
sampling probe is now in the correct height.
Note
The procedure described below assumes that you are using a spectrometer
operated via ICP Syngistix software. If you are using an AA or ICP-MS
instrument, please refer to the information on operating an autosampler in
the on-line help for the respective instrument.
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......