249
Removing and Replacing the Axial Flat Optic
3. Slowly and carefully twist out towards the front of the instrument the Axial
Heat Sink Cooling Tube.
4. Slowly and carefully twist in the new Axial Heat Sink Cooling Tube into place.
Removing and Replacing the Axial Flat Optic
You may need to remove and replace the Axial Flat Optic if you notice reduced
Axial sensitivity or counts.
Tools Needed:
•
6mm Hex Mini Ratchet
1. Turn of the plasma.
2. Slide the Axial Sink Cooling tube as far as you can to the right. See the
following figure.
If the plasma has been on, the torch glass will be very hot and can
cause serious burns.
Wait five minutes after turning off the plasma
before you begin these maintenance procedures.
Avertissement: Si le plasma a été mis sur le verre de la torche sera
très chaude et peut causer des brûlures graves.
Attendez cinq minutes
après avoir éteint le plasma avant de commencer ces procédures
d'entretien.
!
Warning
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......