291
Peristaltic Pump
Installing the Drain Tubing
Install the drain tubing on an open channel of the peristaltic pump as follows:
1. Pre-stretch a piece of 1.14 mm (0.045 in.) pump drain tubing.
2. Carefully stretch the new tubing around the pump head and place the clips on
the pump tubing into the slots, using the slots in the same position on each side
of the pump head. Be careful not to twist or kink the pump tubing during this
process.
3. Attach the Teflon tubing to the pump tubing for the drain.
Make sure you install the tubing on the pump so that the waste liquid is
pumped
out
of the spray chamber. See Figure 5-40.
If the tubing is installed
in the wrong direction, the spray chamber will flood.
4. For the tubing that leads to the drain bottle, use one piece of 1.07 mm i.d.
(0.042 in.) Teflon tubing (Part No. 09985708). Bevel one end by cutting it at
an angle.
5. Insert the Teflon tubing into the other side of the pump tubing. Lead this tubing
to the drain bottle. Refer to Figure 5-40.
6. Adjust the tubing as described in the next procedure.
3
Sleeve (Part No. N0695415)
8
Teflon Tubing 1 mm i.d.
(Part No. 09985708)
4
Spray Chamber
9
Drain bottle
5
PVC Tubing 1.5 mm i.d.
(Part No. 09985735)
10
Peristaltic Pump
Item
Description
Item
Description
Summary of Contents for AVIO 500
Page 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Page 2: ......
Page 3: ...Avio 500 Customer Hardware and Service Guide ...
Page 12: ...Contents 10 ...
Page 30: ...28 ...
Page 31: ...C h a p t e r 1 ...
Page 32: ......
Page 33: ...Safety Practices 1 ...
Page 58: ...56 Safety Practices ...
Page 59: ...C h a p t e r 2 ...
Page 60: ......
Page 76: ...74 Preparing Your Laboratory ...
Page 77: ...C h a p t e r 3 ...
Page 78: ......
Page 79: ...System Description 3 ...
Page 115: ...C h a p t e r 4 ...
Page 116: ......
Page 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Page 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Page 196: ...194 Installation ...
Page 197: ...C h a p t e r 5 ...
Page 198: ......
Page 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Page 316: ...314 Maintenance ...
Page 317: ...C h a p t e r 6 ...
Page 318: ......
Page 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Page 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Page 346: ...344 Troubleshooting ...
Page 354: ......