Command Syntax
STATus:QUEStionable[:EVENt]?
Returned Format
<decimal data> = the sum (between 0 and 16384) of all bits that are true. See the table for
STATus:QUEStionable:CONDition
6.3.6.7
STATus:PRESet
Function
Enables Device Status Reporting. This command has an SCPI standardized effect on the
status data structures. The purpose is to precondition these toward reporting only device-
dependent status data.
It only affects enable registers. It does not change event and condition registers.
The IEEE-488.2 enable registers, which are handled with the common commands
*SRE and *ESE remain unchanged.
The command sets or clears all other enable registers. Those relevant for this device
are as follows:
It sets all bits of the Device status Enable Registers to 1.
It sets all bits of the Questionable Data Status Enable Registers and the Operation
Status Enable Registers to 0.
The following registers never change in the GSG-5x, but they do conform to the
standard :STATus:PRESet values.
All bits in the positive transition filters of Questionable Data and Operation status
registers are 1.
All bits in the negative transition filters of Questionable Data and Operation status
registers are 0.
Command Syntax
STATus:PRESet
326
User Manual GSG-5/6 Series Rev. 27
6.3 Command Reference
Summary of Contents for GSG-5 Series