Command Syntax
SOURce:ONECHN:LOSDynamics:SETtings?
Example
SEND:
SOURce:ONECHN:LOSD:SET?
READ:
0.005,0.1,20,2
6.3.3.37 SOURce:ONECHN:LOSDynamics:CONTrol
Function
Starts, restarts or stops the line of sight dynamics profile. This command can only be used
while RF is generated. Before starting the profile its parameters must be set using the
SOURce:ONECHN:LOSDynamics:SETtings
command.
After the profile is started, the Doppler frequency shift is controlled automatically and can
neither be changed using the
SOURce:ONECHN:FREQuency
command, nor via the front
panel.
Command Syntax
SOURce:ONECHN:LOSDynamics:CONTrol <START|STOP>
Parameters
<START>
– starts line of sight dynamics profile if the profile is not running or restarts the
profile applying new parameters
<STOP>
– stops line of sight dynamics profile
Example
SOUR:ONECHN:LOSD:CONT START
6.3 Command Reference
User Manual GSG-5/6 Series Rev. 27
261