Are you sure you want to restore factory defaults?
Confirmation request before restoring factory defaults.
IP address configuration failed.
Configuring the Internet Protocol address failed. Please check your network settings and
try again.
Cannot save unit parameters.
Saving of settings failed. Try restarting the device and saving parameters again. If it still
fails, then please contact service.
Cannot load unit parameters.
The device was not able to use the configured parameters. Defaults are used in this case.
Please go to the Menu and set the parameters again and store them with EXIT button.
Cannot load unit calibration data.
Unit is not calibrated, or calibration data is corrupted. The device should be re-calibrated.
Please contact service. Device can be used, but the observed power levels may differ from
the ones shown in display.
Cannot save unit calibration data.
Manual calibration data could not be saved. Restart unit and try again. If saving is still not
possible, contact service.
Problem with the license file.
The License file is corrupted. The device will work, but only as GSG-51. Contact service.
Are you sure you want to calibrate the unit?
Confirmation for user calibration.
Password is invalid.
Calibration password entered is invalid. Try again.
Save calibration?
After manual calibration you can choose to save or not save the values.
178
User Manual GSG-5/6 Series Rev. 27
5.2 Messages
Summary of Contents for GSG-5 Series
Page 2: ......
Page 4: ...Blank page II User Manual GSG 5 6 Series...
Page 42: ...BLANK PAGE 2 5 Signal Power Level Considerations 26 User Manual GSG 5 6 Series Rev 27...
Page 382: ...BLANK PAGE 6 7 Revision History SCPI Guide 366 User Manual GSG 5 6 Series Rev 27...