![NF ZM2371 Instruction Manual Download Page 68](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900068.webp)
3.5 Basic Operations
3-38
ZM2371/ZM2372
3.5.6.4 Trigger
The
ZM2371 / ZM2372
receive a trigger signal to start the measurement.
With the trigger setting menu, set the trigger source and measurement sequence.
( a ) Continuous drive
( b ) Triggered drive
( c ) When measuring DC
resistance
Figure 3
–5 Measurement sequence
Trigger signal
Measurement
signal output
Signal
acquisition
Parameter
calculation
Measurement
result
Trigger delay time
Trigger delay time
Trigger delay time
Rdc
Z