8.7 Performance Testing
8-4
ZM2371/ZM2372
8.7 Performance Testing
Performance testing is conducted as part of preventive maintenance to prevent performance
degradation of the
ZM2371/ZM2372
. Besides, conduct it if needed after acceptance inspection,
periodic inspection or repair.
If the result of a performance testing does not meet the specifications, calibration or repair
is required. Contact NF Corporation or one of our representatives.
The performance testing should be conducted in the following conditions.
Power voltage:
100 to 230V ±10% (250V or less)
Ambient temperature: 23±5°C
Ambient humidity:
20 to 70%RH, non-condensing
Warm up:
30 min or more
Operation mode:
0: Set it with refer to
“6.1 Outline and Switching of Operation
Mode”
.
Take the following precautions when you conduct a performance testing.
The setting contents for each test item contains the descriptions of items which should be
further changed after initializing the setting.
Setting the initialize
“3.5.4 Initialization”
_
SHIFT
_
+ [INIT],
_
1
_
Key operation
8.7.1 Measurement Frequency Accuracy
Connection:
H
CUR
terminal → Frequency counter input
Setting:
After the initializing operation (
_
SHIFT
_
+ [INIT],
_
1
_
Key operation), set the
measurement frequency 1kHz (initial value) and measurement signal level
1Vrms (initial value).
Measurement: Measure the frequency with the frequency counter.
Evaluation:
If the values on the counter are within the preset value
±
100ppm
(0.999900kHz to 1.000100kHz), it is normal.
8.7.2 Measurement Signal Level Accuracy
Connection:
H
CUR
terminal → AC voltmeter input
The cable length should be approx 1m or less.
Setting:
After initializing the setting, set the measurement frequency and
measurement signal level according to the contents in table stated below.
Measurement:
Measure the output voltage with the AC voltmeter.
Evaluation:
If the values indicated on AC voltmeter are within the specification range in
table, it is normal.
Measurement frequency
120Hz
1kHz
10kHz
100kHz
Values indicated on AC voltmeter
Specification
:
±
(10% + 5mVrms)
Measurement
signal level
0.1Vrms
85mVrms to 115mVrms
1Vrms
0.895 Vrms to 1.105 Vrms
5Vrms
4.495Vrms to 5.505 Vrms