![NF ZM2371 Instruction Manual Download Page 62](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900062.webp)
3.5 Basic Operations
3-32
ZM2371/ZM2372
3.5.6.2 Measurement Signal Level
The measurement signal level is displayed at the lower right of measurement screen.
・ ・ ・ ・ ・ ・ ・
1 . 0 0 0 0 k
Measurement frequency Hz
・ ・ ・ ・ ・ ・ ・
1 . 0 0 V
c
Measurement signal level
Vrms
_
LEVEL
_
Press the
_
LEVEL
_
key to display the measurement signal level setting menu.
L e v e l : 1 . 0 0 V
Current setting
0 . 0 1 0 V t o 5 . 0 0 V
Range of settable values
The setting resolution is 3 digits (1mV when < 100mV).
Though the value is a root-mean-square value (unit: Vrms), it is simply displayed as V.
If a setting is made or the EXIT operation is performed, the measurement screen comes
back.
_
LEVEL
_
For the direct-current resistance Rdc, the voltage and current are measured with two
measurement signals of about +1.4V and -1.4V, and Rdc is obtained from a difference of
them, regardless of a setting of measurement signal level.