5.6 Command Explanation
5-39
ZM2371/ZM2372
:CALCulate2:FORMat {Q|D|PHASe|X|B|RS|RP|G|LP|RDC|IMAGinary|REAL}
:CALCulate2:FORMat?
Description:
Sets the secondary parameters to be measured.
Parameters:
Q
Quarity factor
(=
1/D
)
D
Dissipation factor (= tan
)
PHASe
Phase angle of impedance (
, unit: degree)
X
Reactance Xs expressed with series equivalent circuit
(unit:
)
B
Susceptance Bp expressed with parallel equivalent circuit
(unit: siemens)
RS
Resistance Rs expressed with series equivalent circuit
(unit:
)
RP
Resistance Rp expressed with parallel equivalent circuit
(unit:
)
G
Conductance Gp expressed with parallel equivalent circuit
(unit: siemens)
LP
Inductance Lp expressed with parallel equivalent circuit
(unit:
)
RDC
DC resistance (unit:
)
IMAGinary Imaginary part of immittance (X or B)
REAL
Real part of immittance (either Rs, Gp or Rdc)
Note: Immittance is generic name of impedance and
admittance.
Setting example:
:CALC2:FORM D
Sets the secondary parameter to D.
Response:
{Q|D|PHAS|X|B|RS|RP|G|LP|RDC|IMAG|REAL}
Query example:
:CALC2:FORM?
Response example: D
The secondary parameter is dissipation factor D.
Remarks:
Equivalent circuit assumed in the measurement of secondary
parameters is as follows:
PHASe, D, Q, RDC
Not depend on the equivalent circuit.
X, RS
Series equivalent circuit
G, B, LP, RP
Parallel equivalent circuit
These parameters do not depend on [:SENSe]:FUNCtion[:ON]
command and equivalent circuit setting.
IMAGinary, REAL
These parameters depend on [:SENSe]:FUNCtion[:ON] command
and equivalent circuit setting.
See the following table
[:SENSe]:FUNCtion[:ON]
IMAGinary
REAL
FIMPedance (Series equivalent circuit)
X
Rs
(=
ESR
)
FADMittance (Parallel equivalent circuit) B
G
FIMPedance,FRESistance
X
Rdc
FADMittance,FRESistance
B
Rdc
The setting of equivalent circuit (series/parallel) and the setting of
automatic selection of equivalent circuit do not affect the secondary
parameter setting. However, the secondary parameters actually
measured and displayed may vary since [:SENSe]:FUNCtion[:ON]
command and equivalent circuit setting are interlocked.