5.7 Status System
5-80
ZM2371/ZM2372
Table 5
–6 Contents of operation condition register and event register
Bit
Weight
Content of condition register
(“1” when specified status is fulfilled)
Content of event register
(Condition for setting to “1”)
-
-
-
15
14
13
32768
16384
8192
Always at 0 (unused)
Always at 0
-
12
4096 Always at 0 (unused)
Always at 0
-
11
2048 Always at 0 (unused)
Always at 0
BUF3 10
1024 BUF3 is full
BUF3 becomes full
BUF2
9
512
BUF2 is full
BUF2 becomes full
BUF1
8
256
BUF1 is full
BUF1 becomes full
CORR
7
128
Either OPEN, SHORT, or LOAD
correction is being measured
Correction measurement
completed
WARM 6
64
Always at 0 (unused)
Always at 0
WTRG
5
32
Trigger waiting
Device waits for trigger
MEAS
4
16
Measuring (/EOM signal of
handler interface is on high level)
Measurement completed
SWE
3
8
Signal acquiring (/INDEX signal
of handler interfaced is on high
level)
Signal acquisition completed
RANG
2
4
Range switching by automatic
range selection function
Measurement range has been
established
SETT
1
2
Signal settling
(trigger is being delayed)
Signal has settled
(trigger delay finished)
-
0
1
Always at 0 (unused)
Always at 0