9.1 Specifications
9-3
ZM2371/ZM2372
Trigger source
INT
Internal (automatic continuous trigger)
MAN
Manual
EXT
Handler interface
BUS
Remote control
Trigger delay time
Setting range 0.000s to 999.999s, Resolution 0.001s
(Time after input of trigger until start of signal acquisition)
Triggered drive
Drive only at measurement / Continuous drive selectable
(Measurement signal can be output only during the time from trigger
to completion of signal acquisition)
Measurement speed RAPid / FAST / MEDium / SLOW / VerySLOw
Typical measurement time
(Supplementary value. Time from input of trigger to output of
measurement end signal EOM)
Measurement
frequency
RAP
FAST
MED
SLOW
VSLO
120Hz
10ms
10ms
26ms
126ms
501ms
1kHz
2ms
5ms
25ms
121ms
501ms
10kHz
3ms
5ms
25ms
122ms
502ms
100kHz
3ms
5ms
25ms
122ms
502ms
Conditions: Measurement range fixed, Trigger delay time = 0,
Averaging count = 1, Secondary parameter≠Rdc
Signal acquisition time is the value subtracted by about 1ms from the
above value.
DUT can be replaced immediately after the completion of signal
acquisition. After replacement of DUT, the signal settling time is
required additionally.
Additional time when measuring direct-current resistance Rdc
(supplementary vaue)
RAP
FAST
MED
SLOW
VSLO
(DC)
148ms
148ms
148ms
215ms
613ms
Conditions: DC resistance measurement range fixed,
Trigger delay time = 0, Averaging count = 1
Measurement range
Measurement range Recommended range
Measurable range
1M
1M
to 11M
≧
900k
100k
100k
to 1.1M
≧
90k
10k
10k
to 110k
≧
9k
1k
1k
to 11k
≧
0.9k
100
9
to 1.1k
No limitation
10
0.9
to 10
≦
11
1
90m
to 1
≦
1.1
100m
9m
to 100m
≦
110m
Measurable range:
Approximate range in which measurement and
display are possible (supplementary value).
Recommended range: Recommended operating range for high
accuracy measurement.
Limitation of measurement range
When frequency > 20kHz, 1M
range cannot be used.
When minimum output impedance setting is 100
, 1
range or
100m
range cannot be used.
Measurement range Auto / Manual
selection