![NF ZM2371 Instruction Manual Download Page 240](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900240.webp)
9.1 Specifications
9-5
ZM2371/ZM2372
Kz: Residual impedance coefficient
Frequency
Kz [%]
DC (0Hz), Frequency
≦
120Hz
(0.003 + Kc) / Zx[
]
120Hz < Frequency
≦
1kHz
(0.005 + Kc) / Zx[
]
1kHz < Frequency
≦
10kHz
(0.005 + 0.002 × Frequency [kHz]+ Kc) / Zx[
]
10kHz < Frequency
≦
100kHz
(0.0025 × Frequency [kHz] + Kc) / Zx[
]
Cable length coefficient Kc = 0.001
Fequency [kHz]
(Cable length [m])
2
Use the next table when all of two conditions listed below are satisfied. In this case,
unless Az exceeds 10 [%], the measurement accuracy is not a supplementary value
but a guaranteed value.
Measurement range Zx is 10
.
Impedance magnitude|Z| is less than 0.45
.
Frequency
Kz [%]
DC (0Hz), Frequency
≦
10kHz
(0.05 + Kc) / Zx[
]
10kHz < Frequency
≦
100kHz
(0.1 + 0.002 × Frequency [kHz] + Kc) / Zx[
]
Ky: Residual admittance coefficient
Frequency
Ky [%]
DC, Frequency
≦
120Hz
Zx[
] / ( 3 × 10
8
)
120Hz < Frequency
≦
100kHz
Zx[
] × Frequency [kHz] / ( 3 × 10
7
)
K
T
: Temperature-dependent coefficient
Ambient temperature (T °C)
K
T
0 to +18
1 + 0.1 × (18-T)
+18 to +28
1
+28 to +40
1 + 0.1 × (T-28)
K
B
: DC bias coefficient
Internal DC
bias
Measurement
range Zr
KB [%]
Frequency
≦
1kHz
1kHz <
Frequency
≦
10kHz
Frequency
>10kHz
Disabled
All ranges
0
0
0
Enabled
(*1)
1M
0.005
0.02
0.02
100k
0.002
0.003
0.01
100
to 10k
0.001
0.002
0.01
10
0.01
0.01
0.02
100m
, 1
0.05
0.1
0.2
*1: when open compensation and short compensation are performed at the conditions
of internal DC bias enabled and the bias voltage 0V
At all times, K
B
= 0 for the direct-current resistance Rdc.