3.5 Basic Operations
3-30
ZM2371/ZM2372
Negative display of Q, D, R, G
When the phase angle of impedance
> +90
or
< -90
due to a measurement error, the
loss angle seems to be a negative value, and thus the Q, D, Rs (ESR), Rp, and G become
negative values. Also, Rdc may become a negative value due to a measurement error of
voltage or current.
Parameters that can be selected by remote control only
The following options can be selected by remote control only. They cannot be selected from
the panel.
For details
Chapter 5 :CALCulate2: FORMat command
IMAG
Imaginary part of immittance (X or B)
REAL
Real part of immittance (either of Rs, G, or Rdc)
Immittance is generic name of impedance and admittance.