5.6 Command Explanation
5-70
ZM2371/ZM2372
:SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet <DC offset>
:SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet?
Description:
Sets/queries internal DC offset(/bias) voltage superimposed on the
measurement signal.
To actually superimpose the DC offset, set additionally the following
command so as to superimpose.
・・・
:SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet:STATe
Parameters:
<DC offset> {numeric value, range 0 to 2.50, resolution 0.01, unit: V}
Unit V and MAX / MIN can be used. Example: 0.5V (= 0.5)
Setting example:
:SOUR:VOLT:OFFS 1.5
Sets the DC offset voltage to1.5.
Response:
<DC offset> {numeric value, format NR3}
Query example:
:SOUR:VOLT
:
OFFS?
Response example: +1.00
The DC offset voltage is +1.00V(1.5V).
:SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet:STATe {ON|OFF|1|0}
:SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet:STATe?
Description:
Sets whether internal DC offset(/bias) voltage is superimposed on the
measurement signal.
Parameters:
{ON|1}
DC offset voltage is superimposed.
{OFF|0}
DC offset voltage is not superimposed.
The DC offset cannot be turned on when the triggered drive is enabled.
Setting example:
:SOUR:VOLT:OFFS:STAT 1
Sets so as to superimpose the DC offset voltage.
Response:
{1|0}
Query example:
:SOUR:VOLT
:
OFFS:STAT?
Response example: 1
DC offset voltage is superimposed.
:SOURce:VOLTage:MODE {CONTinuous|SYNChronous}
:SOURce:VOLTage:MODE?
Description:
Sets/queries the triggered drive.
Parameters:
CONTinuous
Triggered drive is disabled and measurement signal is
output continuously.
SYNChronous
Triggered drive is enabled and measurement signal is
output only during measurement. DC offset is turned
off.
Setting example:
:SOUR:VOLT:MODE SYNC
Sets so as to output the measurement signal only during measurement
in synchronization with the trigger.
Response:
{CONT|SYNC}
Query example:
:SOUR:VOLT:MODE?
Response example: SYNC
The measurement signal is output only during measurement.