5.8 Trigger System
5-82
ZM2371/ZM2372
5.8 Trigger System
The trigger system of
ZM2371 / ZM2372
is shown below.
(:INIT:CONT ON state)
(:INIT:CONT OFF state)
Power ON
_
SHIFT
_
+ [INIT],
_
1
_
key
Full initialization
*RST
Operation mode setting
Idle
:ABORT
:INIT:CONT ON state
or
:INIT command
Trigger waiting
:READ?
Trigger signal (either of the followings)
1) :TRIG:SOUR INT
2) In :TRIG:SOUR MAN, TRIG key on
the panel
3) In :TRIG:SOUR EXT, TRIG signal of
handler interface
4) In :TRIG:SOUR BUS, *TRG
、
GET
5) :TRIG (when trigger source is other
than INT)
Trigger delay
(:TRIG:DEL)
Measurement once
Figure 5
–5 Trigger system