:CALCulate1:LIMit:UPPer:STATe {ON|OFF|1|0}
:CALCulate1:LIMit:UPPer:STATe?
Description:
Sets/queries whether the upper limit value is compared or not in the
limit comparison of primary parameters.
Parameters:
{ON|1}
Upper limit value is compared.
{OFF|0}
Upper limit value is not compared (corresponding to No
Limit of panel operation).
Setting example:
:CALC1:LIM:UPP:STAT ON
Sets so that the upper limit value is compared in the primary parameter
limit comparison.
Response:
{1|0}
Query example:
:CALC1:LIM:UPP:STAT?
Response example: 1
The upper limit value is compared in the primary parameter limit
comparison.
:CALCulate1:MATH:EXPRession:NAME {DEV|PCNT}
:CALCulate1:MATH:EXPRession:NAME?
Description:
Sets/queries the deviation display format of primary parameters.
To display the deviation, set the deviation display format, and further
set so that the deviation is displayed using :CALCulate1:MATH:STATe
ON command.
Set the referenced value with :DATA[:DATA] command.
Parameters:
DEV
Deviation: Deviation from reference value (= measured
value
reference value) is displayed.
PCNT
Deviation %: Deviation (to reference value) is displayed with
percent value.
Setting example:
:CALC1:MATH:EXPR:NAME PCNT
Sets the deviation display format of primary parameters to deviation %.
Response:
{DEV|PCNT}
Query example:
:CALC1:MATH:EXPR:NAME?
Response example: PCNT
The deviation display format of primary parameters is deviation %.
:CALCulate1:MATH:STATe {ON|OFF|1|0}
:CALCulate1:MATH:STATe?
Description:
Sets/queries whether the primary parameter deviation display / output
is executed or not.
Parameters:
{ON|1}
Enables the deviation display / output.
{OFF|0}
Disables the deviation display / output.
Setting example:
:CALC1:MATH:STAT ON
Sets to enables the primary parameter deviation display / output.
Response:
{1|0}
Query example:
:CALC1:MATH:STAT?
Response example: 1
The primary parameter deviation display / output is enabled.
Remarks:
If the primary parameters or secondary parameters are changed
with :CALCulate{1|2}:FORMat command, the deviation display / output
of both primary and secondary parameters is disabled automatically.
Set the primary parameters and secondary parameters prior to the
deviation display / output.
Summary of Contents for ZM2371
Page 1: ...NF Corporation LCR METER ZM2371 ZM2372 Instruction Manual...
Page 2: ......
Page 3: ...LCR METER ZM2371 ZM2372 Instruction Manual DA00026021 007...
Page 250: ......
Page 252: ......
Page 253: ......
Page 255: ...BLANK...