9.1 Specifications
9-2
ZM2371/ZM2372
9.1 Specifications
Measurement parameters
Primary parameters
|Z|, |Y|, L, C, R, G
For equivalent circuit of L, C, and R, Parallel / Series / Auto
Selection are selectable.
Secondary parameters
Q, D,
, X, B, Rs, Rp, G, Lp, Rdc (direct-current resistance)
Auto parameter selection Primary parameters (including equivalent circuit) and secondary
parameters can be selected automatically.
Measured value display range
|Z|
0.000m
to 999.999M
R (Rs, Rp, Rdc), X
0
,
(0.001m
to 999.999M
)
|Y|
0.00nS to 9.99999kS
G, B
0S,
(0.01nS to 9.99999kS)
C (Cp, Cs)
0F,
(0.00001pF to 999.999kF)
ex. 0.000pF to 999.999mF (1kHz)
0.00pF to 9.99999F (120Hz)
L (Ls, Lp)
0H,
(0.001nH to 99.9999GH)
ex. 0.0000
H to 99.9999kH (1kHz)
Q, D
0,
(0.00001 to 99999.9)
±180.000°
Actual measurement and display ranges of respective
parameters are restricted by the measurement range or
frequency.
Measurement conditions
Measurement frequency
1mHz to 100kHz, Resolution 5 digits (1mHz when < 10Hz),
0.01%
Measurement signal level 10mV to 5.00V, Resolution 3 digits (1mV when < 100mV),
(10% + 5mV)
RMS value when output is open
ALC
Constant voltage drive / Constant current drive / Invalid
Voltage setting range 10mV to 5.00V, Setting resolution 3 digits
(1mV when < 100mV)
Current setting range 1
A to 200mA, Setting resolution 3 digits
(0.1
A when < 10
A)
The constant control range will be narrower than the above
specifications depending on the product dispersion or DUT’s
impedance.
The current range is restricted depending on the measurement
range.
Output impedance
5
/ 25
/ 100
(supplementary values). Automatically selected
according to the measurement range.
Minimum output impedance restriction function is available.
5
is selected automatically only when the following conditions
are satisfied:
Minimum output impedance setting = 5
,
Measurement range
≦
10
,
Measurement signal level
≦
1V, ALC = invalid,
Internal DC bias = Off, Secondary parameter ≠ Rdc
Internal DC bias
0V to +2.50V, Resolution 0.01V,
(5% +3mV) when output is open
On/Off is possible