![NF ZM2371 Instruction Manual Download Page 204](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900204.webp)
5.7 Status System
5-77
ZM2371/ZM2372
5.7.3 Standard Event Status
The standard event status structure is shown on
Fgure 5-3
. Details about status are shown
in
Table 5-5
. When the standard event status enable register bits are set to 1, the
corresponding standard event status register bits become valid. When at least one such bit
is set to 1, the status bit register ESB bit is set to 1.
Figure 5
–3 Standard event status structure
Table 5
–5 Standard event status register contents
Bit
Weigh
t
Contents
PON
7
128
Power on
Set to 1 when the power is on. If set to 0 by a register readout,
remains at 0 until the power is turned on again.
URQ
6
64
User request
Always at 0 (unused)
CME
5
32
Command error
Set to 1 when there is a syntax error in program code.
EXE
4
16
Execution error
Set to 1 when parameters are set beyond possible range values or
when settings are contradictory.
DDE
3
8
Device dependent error
Set to “1” when the error queue overflows.
QYE
2
4
Query error
Set to 1 when a readout has been attempted on an empty response
message output buffer or when the response message output buffer
data has been lost.
RQC
1
2
Request control
Always at 0 (unused)
OPC
0
1
Operation complete
Set to “1” when a processing of all commands up to *OPC command
completed.
Power on PON
Always at 0 (unused) URQ
Command error CME
Execution error EXE
Always at 0 (unused) DDE
Query error QYE
Always at 0 (unused) RQC
Operation complete (OPC)
7
6
5
4
3
2
1
0
7
6
5
4
3
2
1
0
L
og
ic
al
O
R
(Standard event status register) ESR
(Standard event status enable register) ESE
Standard event status
summary
Status byte ESB (bit5)
&: logical AND
&