5.6 Command Explanation
5-55
ZM2371/ZM2372
[:SENSe]:AVERage:COUNt <count>
[:SENSe]:AVERage:COUNt?
Description:
Sets/queries the measurement averaging count.
The signal acquisition time is averaging-count multiples of the base
value determined by the measurement speed.
To execute the averaging, enable the averaging function by
[:SENSe]:AVERage[:STATe] command additionally.
Parameters:
<count> {numeric value, range 1 - 256} MAX / MIN can be used.
Setting example:
:AVER:COUN 100
Sets the averaging count to 100.
Response:
{numeric value, format NR1}
Query example:
:AVER:COUN?
Response example: +100
The averaging count is 100.
[:SENSe]:AVERage[:STATe] {ON|OFF|1|0}
[:SENSe]:AVERage[:STATe]?
Description:
Sets/queries whether the averaging is executed or not.
Set the averaging count additionally with [:SENSe]:AVERage:COUNt
command.
Parameters:
{ON|1}
Enables the averaging function.
{OFF|0}
Disables the averaging function.
Setting example:
:AVER ON
Sets so as to enable the averaging function.
Response:
{1|0}
Query example:
:AVER?
Response example: 1
The averaging function is enabled.
Remarks:
If the averaging count is set to 2 or larger value by the panel operation,
the averaging function is automatically enabled. If the averaging count
is set to 1 by the panel operation, the averaging function is
automatically disabled.