4.4 Connecting to the Part handler (Handler Interface)
4-21
ZM2371/ZM2372
Checking the operation of handler interface
You can check the operation through outputting the given signal into handler interface or
monitoring the input signal.
Handler interface check menu is as follows:
H a n d l e r C h e c k T R I G t R C L s s s V L D v L O C K k
Input signal monitor
0 ) O F F 1 ) D O W N 2 ) U P 3 ) E O M B I N n n
Output options
Signal name which is in dummy output
Perform the EXIT operation to return to the measurement screen.
Input signal monitor
Numeric values appear while the status of Input signal line is being regarded as the
negative logic input.
Signal name
Display
name
Status
Descriptions of status
/TRIG
TRIG
t
0: high level, 1: low level
(It is regarded as negative logic regardless of the setting
of trigger polarity)
/RCL0../RCL4
RCL
sss
0..127: next 7-bit is converted into decimal 3-digit before
displaying
From high order, Pin 29, Pin 4, /RCL4 (MSB), /RCL0
(LSB)
/RCL-VALID
VLD
v
0: High level, 1: Low level
/KEY_LOCK
LOCK
k
0: High level, 1: Low level
Dummy output signal
Select the each option with the numeric key, the following operation starts:
The signal name (abbreviation) being output appears on the lower right of display.
0)OFF
Allows all outputs to turn off (high level)
When handler interface checking menu appears, this status comes out.
1)DOWN Allows output signal to decrease by one smaller in Bin number (see below)
2)UP
Allows output signal to increase by one larger in Bin number (see below)
Example of UP operation (DOWN operation allows reversed movement)
(OFF)
→
/BIN1
→
/BIN2
→・・・
/BIN14
→
/OUT OF BINS
→
/S-NG
→
/INDEX
→
/EOM
→
/ERR
→
(OFF)
Only one of output signals is ON (low level), but others are OFF (high level).
However, all output signals are OFF at (OFF) position.
3)EOM
/INDEX and /EOM are operated as is the case with the measurement when
trigger is applied but it may not affect on other signals. The change of
signals with the times is fixed as below.
Approx 10ms
/INDEX
/EOM
Approx 1ms
Approx 1ms
Figure 4
–5 Dummy output timing