![NF ZM2371 Instruction Manual Download Page 52](http://html1.mh-extra.com/html/nf/zm2371/zm2371_instruction-manual_1684900052.webp)
3.5 Basic Operations
3-22
ZM2371/ZM2372
Selection of measurement parameters
Selection of primary parameter
Press the
_
Pri
_
key to display the primary parameter selection menu (see below), and select
a primary parameter with a numeric key. Also, the measurement parameters (primary
and secondary) can be selected automatically.
For details
“3.5.5.1 Selection of Primary Parameter”
P r i m a r y pa r a m e t e r : C
A u t o : O F F
Current setting
0)Auto OFF 1)ON 2)L 3)C 4)R 5)Z 6)Y 7)G
Options
If either L, C, or R is selected, the equivalent circuit selection menu is displayed.
Selection of primary parameter equivalent circuit
For the L, C, and R, specify the equivalent circuit additionally.
Press the
_
CKT
_
key to display the equivalent circuit selection menu (see below), and
select Parallel (suffix p) or Series (suffix s) with a numeric key. Automatic selection is also
possible.
For details
“3.5.5.2 Setting the Equivalent Circuit”
E q u i v a l e n t C K T : S e r i e s
A u t o : O F F
Current setting
0 ) A u t o O F F 1 ) O N 2 ) P a r a l l e l 3 ) S e r i e s
Options
Selection of secondary parameter
Press the
_
Sec
_
key to display the secondary parameter selection menu (see below), and
select a secondary parameter with a numeric key. The options can be switched to those on
the second page with the
_
EXP
_
| [NEXT]
key.
For details
“3.5.5.3 Setting the Secondary Parameter”
S e c o n d a r y pa r a m e t e r : D
Current setting
0)Q 1)D 2)
3)X 4)B 5)Rs
>NEXT
Options (First page)
6)Rp 7)G 8)Lp 9)Rdc
<PREV
Options (Second page)
_
CKT
_
_
Pri
_
_
Sec
_
Summary of Contents for ZM2371