5.6 Command Explanation
5-35
ZM2371/ZM2372
:CALCulate1:FORMat {Z|Y|R|RP|RS|G|C|CP|CS|L|LP|LS|REAL|MLINear}
:CALCulate1:FORMat?
Description:
Sets/queries the primary parameter (and equivalent circuit) to be
measured.
Parameters:
Z
Magnitude of impedance |Z| (unit: ohm
)
Y
Magnitude of admittance |Z| (unit: S siemens)
R
Resistance Rp or Rs (unit:
)
RP
Resistance Rp expressed with parallel equivalent circuit
(unit:
)
RS
Resistance Rs expressed with series equivalent circuit
(unit:
)
G
Conductance GP expressed with parallel equivalent circuit
(unit: S)
C
Capacitance Cp or Cs (unit: F farad)
CP
Capacitance Cp expressed with parallel equivalent circuit
(unit: F farad)
CS
Capacitance Cs expressed with series equivalent circuit
(unit: F farad)
L
Inductance Lp or Ls (unit: H henry)
LP
Inductance Lp expressed with parallel equivalent circuit
(unit: H henry)
LS
Inductance LS expressed with series equivalent circuit
(unit: H henry)
REAL
Real part of immittance (either Rs or Gp)
MLINear Magnitude of immittance (|Z|or |Y|)
Note: Immittance is generic name of impedance and admittance.
Setting example:
:CALC1:FORM CS
Sets the primary parameter to the capacitance expressed with series
equivalent circuit.
Response:
{Z|Y|R|RP|RS|G|C|CP|CS|L|LP|LS|REAL|MLIN}
Query example:
:CALC1:FORM?
Response example: CS
The primary parameter is the capacitance expressed with series equivalent
circuit.
Remarks:
Equivalent circuit assumed in the measurement of primary parameters is
as follows:
Z, Y
Not depend on the equivalent circuit.
RS, CS, LS
Series equivalent circuit
RP, CP, LP, G
Parallel equivalent circuit
These parameters do not depend on [:SENSe]:FUNCtion[:ON]
command and equivalent circuit setting.
REAL, MLINear, R, C, L
These parameters depend on [:SENSe]:FUNCtion[:ON] command and
equivalent circuit setting.
See below.
<When automatic selection of equivalent circuit is disabled,
(:CALC1:CKIT:AUTO OFF)>
[:SENSe]:FUNCtion[:ON]
REAL MLINear
R
C
L
FIMPedance
(Series equivalent circuit)
Rs
|Z|
Rs Cs Ls
FADMittance
(Parallel equivalent circuit)
G
|Y|
Rp Cp Lp
FIMPedance,FRESistance
Rs
|Z|
Rs Cs Ls
FADMittance,FRESistance
Rp
|Y|
Rp Cp Lp
<When automatic selection of equivalent circuit is enabled,
(:CALC1:CKIT:AUTO ON)>
For R, C, and L, the equivalent circuit automatic selecting function has
priority over the setting of [:SENSe]:FUNCtion[:ON].