5.6 Command Explanation
5-28
ZM2371/ZM2372
5.6.3.2
Subsystem Commands
:ABORt
Description:
Aborts the measurement and places the trigger system in idle state.
When continuous initiation of trigger system is enabled, the device goes
in trigger waiting state automatically. Further when the trigger source
is set to INT (internal), a new trigger is applied automatically to start
the measurement.
The measurement of OPEN, SHORT, and LOAD correction values is
aborted.
Setting example:
:ABOR
:CALCulate:COMParator:AUXBin {ON|OFF|1|0}
:CALCulate:COMParator:AUXBin?
Description:
Sets/queries whether S-NG (primary parameter is within the range but
secondary parameter is out of range) is used as auxiliary bins
(AUX_BIN) independent from OUT OF BINS when the bin sorting is
performed by the comparator function.
Parameters:
{ON|1}
S-NG is used as auxiliary bins.
When judged as S-NG, OUT OF BINS is not output.
{OFF|0}
S-NG is not used as auxiliary bins.
When judged as S-NG, both S-NG and OUT OF BINS are
output. OUT OF BINS includes both primary parameter out
of range and secondary parameter out of range.
Setting example:
:CALC:COMP:AUXB ON
Sets so that S-NG is used as auxiliary bins.
Response:
{1|0}
Query example:
:CALC:COMP:AUXB?
Response example: 1
S-NG is used as auxiliary bins.