Parameter Menus
Operations
2 - 62
0070-10-0666-01
Trio™ Operating Instructions
2.4.6.3
IBP Troubleshooting
MESSAGE/
PROBLEM
REASON
SOLUTION
SENSOR OFF, FAIL
Transducer may be disconnected
during attempt to zero
Check transducer and cables for
connection. If problem persists,
notify hospital technician or
Customer Support
IN DEMO FAIL
Monitor is in DEMO mode during
attempt to zero
Turn monitor OFF. Turn ON to
return to normal monitoring mode
PRESSURE OVER
RANGE, FAIL
Stopcock not open to atmosphere
Turn stopcock to atmosphere prior
to zero attempt. If problem persists,
notify hospital technician or
Customer Support
PULSATILE
PRESSURE, FAIL
Stopcock is not open to atmosphere
and a pulsatile pressure has been
detected
Turn stopcock to atmosphere prior
to zero attempt. If problem persists,
notify hospital technician or
Customer Support
IBP SENSOR OFF
IBP cable may be disconnected from
monitor
Make sure that cable is properly
connected
IBP INIT ERR1
Module failure
Notify hospital technician or
Customer Support
IBP COMM STOP
Module failure or communication
failure
Notify hospital technician or
Customer Support
IBP COMM ERR
Communication error
Notify hospital technician or
Customer Support
IBP ALM LMT ERR
Functional failure
Notify hospital technician or
Customer Support
IBP SYS EXCEED
Systolic pressure value exceeds
measurement range
Check patient, notify physician
IBP DIA EXCEED
Diastolic pressure value exceeds
measurement range
Check patient, notify physician
IBP MAP EXCEED
Mean arterial pressure value exceeds
measurement range
Check patient, notify physician
IBP NEED ZERO-
CAL
Zero calibration must be performed
before measuring IBP
Zero calibration must be
performed. Notify hospital
technician or Customer Support.
Dampened Invasive
Waveform
Air bubbles in tubing
Eliminate air from tubing
Kinked invasive catheter
Change position of catheter, check
patient
Catheter against wall of blood vessel
Check for leaks at connector, flush
catheter
Blood in tubing
Pump pressure bag up to
300 mmHg
Catheter partially occluded
Consult physician
IBP not Displayed /
No Waveform
Cable not plugged in
Check cable connections
Transducer not connected
Check transducer connections
Stopcock turned improperly
Check transducer for proper
alignment
Transducer not zeroed
Check and zero the transducer
Abnormally high or
low readings
Transducer too HIGH or too LOW in
relation to patient's heart
Check patient, adjust transducer,
re-zero