Trio™ Operating Instructions
0070-10-0666-01
2 - 53
Operations
Parameter Menus
2.4.5
Temperature Monitoring
The temperature (TEMP) measurement function of the
Trio
is designed to take continuous
temperature readings from the YSI 400 series probes. One (1) temperature channel is
standard on the
Trio
.
1.
If using a disposable temperature probe:
Connect the preferred disposable temperature probe into the reusable temperature
cable. Plug the cable into the monitor.
2.
If using a reusable temperature probe:
Plug the reusable probe directly into the monitor
3.
Follow probe manufacturer's suggested temperature monitoring sites and suggested
procedures for site checks, reapplication and moving of the probes.
4.
Check the probe site frequently for position, ensuring good contact with the site chosen.
For example, check the skin probe frequently to ensure good skin contact.
NOTE:
A self-test of the temperature measurement is performed
automatically once per hour during the monitoring period.
The test procedure lasts approximately 2 seconds and does
not affect the normal measurement of the temperature
monitoring.
2.4.5.1
Temperature Setup
FIGURE 2-40
Temperature Setup Menu
Accessing the Temperature Setup Menu
To access the
TEMP SETUP
menu, select
TEMP
from the
TEMP
parameter tile using the
Navigator™ Knob. Once in the
TEMP SETUP
menu, use the Navigator Knob to adjust
settings. To close the menu, select
NORMAL SCREEN
(from the menu or the Front Panel
Keypad).