Temperature Setup Menu Selections
Temperature Alarm Limits
Alarms occurring during the process of temperature measurement include two (2) types:
physiological alarms and technical alarms. Physiological alarms occur when the patient’s
temperature values are equal to or exceed set alarm limits. Technical alarms are any
temperature-related alarms, which are not physiological, such as functional failures.
ALM
Allows the user to turn the
TEMP
alarm
ON
or
OFF
. Choose
ON
to
enable the alarm; choose
OFF
to disable the alarm. If the alarm is set
to
OFF
, the alarm
OFF
symbol
will display to the right of
TEMP
on the screen. The
TEMP
alarm is activated when the temperature
value exceeds set high or low
TEMP
values.
ALM PRIORITY
Enables priority selection for the
TEMP
alarm. Choices are
1
,
2
and
3
. Priority 1 alarms are considered the most serious.
ALM PRINT
Enables or disables automatic printing during a
TEMP
alarm
condition. Choose
ON
to enable printing upon
TEMP
alarm. Choose
OFF
to disable printing upon
TEMP
alarm.
TEMP ALM HI
Allows the user to set the upper limit of Temperature 1 alarm. (See
“Temperature Alarm Limits” on page 2-54.)
TEMP ALM LO
Allows the user to set the lower limit of Temperature 1 alarm. (See
“Temperature Alarm Limits” on page 2-54.)
TEMP UNIT
Allows the user to select the unit of measure for temperature. The
selections are
degrees Celsius (°C)
or
degrees Fahrenheit (°F)
.
RESTORE
DEFAULTS
Allows the user to restore the TEMP user default configuration.
FIGURE 2-41
Confirmation Dialog Box
HIGH
ALARM (ºC)
LOW
ALARM (ºC)
HIGH
ALARM (ºF)
LOW
ALARM (ºF)
Temperature (T1)
35 – 43 [40]
26 – 38 [30]
95 – 109.4 [104]
78.8 – 100.4 [86]
*
Factory default values shown in brackets.
*