Trio™ Operating Instructions
0070-10-0666-01
4 - 1
4.0
User Maintenance
4.1
Introduction
This section of the manual outlines routine maintenance to be performed by the user and/or
biomedical technician.
The
Trio
Monitor is designed for stable operation over long periods of time and under
normal circumstances should not require technical maintenance beyond circumstances
described in this section. In general, routine maintenance, calibration and safety checks are
recommended annually, or more often as required by local statutory or hospital
administration practice.
General Maintenance Checks
Before using the monitor, do the following:
1.
Check if there is any mechanical damage;
2.
Check all external cables, inserted modules and accessories for damage;
3.
Check all the functions of the monitor to make sure that the monitor is in good working
condition.
Scheduled Maintenance Checks
1.
Frequently check the device, cables, sensors and wires for damage.
2.
Clean the device as needed.
3.
Perform safety test annually; or once after each disassembly or repair of unit.
4.
Perform all parameter checks and calibration annually or once after each repair of unit.
5.
Test overall function of the device annually.
NOTE:
If any damage is found on the monitor, contact the
biomedical engineer of the hospital or Customer Service
immediately.