Patient Parameter Specifications
Appendix
6 - 10
0070-10-0666-01
Trio™ Operating Instructions
6.2.6.1
Systolic Pressure Measurement
6.2.6.2
Diastolic Pressure Measurement
*Adult and Pediatric
Blood pressure measurements determined with this device are equivalent to those obtained
by a trained observer using the cuff/stethoscope auscultation method, within the limits
prescribed by ANSI/AAMI SP-10:1992,
Electronic or automated sphygmomanometers.
NOTE:
Mean Arterial Pressure (MAP) is defined as:
Mean Pressure 1 = Mean Pressure determined from the
oscillometric profile
Mean Pressure 2 = (2*dia systolic) / 3
Mean Pressure Displayed = (Mean Pressure 1 + Mean
Pressure 2) / 2
6.2.6.3
Static Pressure Measurement
6.2.6.4
Heart Rate from NIBP
Accuracy*:
Mean error is less than ± 5 mmHg, Standard
Deviation is less than ± 8 mmHg
Range:
ADULT MODE
PEDIATRIC MODE
40 to 255 mmHg
40 to 200 mmHg
Accuracy*:
Mean error is less than ± 5 mmHg, Standard
Deviation is less than ± 8 mmHg
Range:
ADULT MODE
PEDIATRIC MODE
10 to 210 mmHg
10 to 150 mmHg
Range:
0 – 325 mmHg
Static Accuracy:
±3 mmHg over the entire range.
Accuracy:
± 2 bpm or 2%, whichever is greater
Resolution:
1 bpm
Range:
ADULT MODE
PEDIATRIC MODE
40 to 240 bpm
40 to 240 bpm