Alarms
Operations
2 - 66
0070-10-0666-01
Trio™ Operating Instructions
HR ALARM SETUP Menu Selections
FIGURE 2-48
HR Alarm Setup Menu
ALM SEL
(Alarm Selection)
Choose
HR ALARM SETUP
.
ALM
NOTE:
In the French configuration, the HR
(Heart Rate) alarm is always ON. It
cannot be disabled by turning it
OFF.
Allows the user to turn the HR (Heart Rate) alarm
ON
or
OFF
.
Choose
ON
to enable the alarm; choose
OFF
to disable the
alarm. If the alarm is set to
OFF
, the alarm
OFF
symbol
will
display to the right of
ECG
on the screen. The HR alarm is
activated when the heart rate is equal to or exceeds set high or
low HR values.
ALM PRIORITY
Allows the user to select the priority of the alarm. Choices are
1
,
2
and
3
. Priority 1 alarms are considered the most serious.
ALM PRINT
Allows the user to enable or disable automatic printing during a
physiological alarm condition. Choose
ON
to enable printing
upon alarm. Choose
OFF
to disable printing upon alarm. See
“Recorder (Optional)” on page 2-82 for details.
HR ALM HI
Allows the user to set the upper limit of the
HR
Rate Alarm Limits” on page 2-27.)
HR ALM LO
Allows the user to set the lower limit of the
HR