Trio™ Operating Instructions
0070-10-0666-01
2 - 85
Operations
Recorder (Optional)
When parameter alarms occur, a maximum of two (2) waveforms can print simultaneously. If
more than two (2) parameter alarms are activated simultaneously, the recorder will print out
those of the highest priority level. If the alarm events are of the same priority level, the most
recent alarm event will be printed. If an alarm event occurs when another alarm printout is
already in progress, the more recent alarm printout will not begin until the current alarm
printout is complete.
Graphic/List Trend Print
To print the graphic trend of the currently selected parameter, select
from the
GRAPHIC TREND
menu. To stop the print job currently in progress, select
on the
front panel keypad.
FIGURE 2-60
Example, Graphic Trend Printout
To print the list trend of all parameters, select
from the
LIST TREND
menu. To stop the
print job currently in progress, select
on the front panel keypad.
FIGURE 2-61
Example, List Trend Printout
Error Log Print
To print information displayed in the
ERROR LOG
menu, select
MAINTENANCE
from the
SYSTEM MENU
. From the
MAINTENANCE
menu, select
ERROR LOG
. Select
from the
ERROR LOG
menu. To stop the print job currently in progress, select the
key on the front panel keypad.
FIGURE 2-62
Example, Error Log Printout