Trio™ Operating Instructions
0070-10-0666-01
6 - 13
Appendix
Patient Parameter Specifications
6.2.9
Temperature Parameter Performance Characteristics
The
Trio
monitor is capable of providing temperature measurements from a single
temperature channel as measured from a YSI 400 probe and meets the requirements of
EN12470-4:2000.
6.2.9.1
Temperature Performance Requirements
6.2.10
SpO
2
Performance Requirements
The
Trio
monitor is capable of providing SpO
2
functional saturation level measurements via
a Masimo SET
®
MS-7 pulse oximeter, a Nellcor
®
MP506 or a Nellcor
®
Nell-3 pulse
oximeter.
6.2.10.1
Agency Requirements:
The SpO
2
function is in accordance with the requirements of EN 865: 1997.
6.2.10.2
Masimo SET
®
SpO
2
Performance Requirements
The Masimo SET MS-7 pulse-oximeter with SET technology is implemented.
Sensor Compatibility:
LNOP
®
or LNCS
®
Series
SpO
2
Display Range: 1 – 100%
SpO
2
Resolution: 1%
Scale:
selectable Celsius or Fahrenheit
Range:
0 °C to 50 °C
32 °F to 122 °F
Resolution:
0.1 °C
0.1 °F
Accuracy:
± 0.1 °C (0 °C to 50 °C), exclusive of probe
errors
± 0.2 °F (32 °F to 122 °F), exclusive of
probe errors
Accuracy inclusive of 400 Series probes:
± 0.2 °C (32 °C – 42 °C)
± 0.3 °C (otherwise)
± 0.4 °F (90 °F – 108 °F)
± 0.6 °F (otherwise)
Probe Excitation: 400 Series:
<100
μ
A, tip to sleeve.