Trio™ Operating Instructions
0070-10-0666-01
2 - 49
Operations
Parameter Menus
MEAN ALM HI
Allows the user to set the upper limit of the Mean
NIBP
alarm. (See
“NIBP Alarm Limits” on page 2-50.)
MEAN ALM LO
Allows the user to set the lower limit of the Mean
NIBP
alarm. (See
“NIBP Alarm Limits” on page 2-50.)
DIA ALM HI
Allows the user to set the upper limit of the Diastolic
NIBP
alarm. (See
“NIBP Alarm Limits” on page 2-50.)
DIA ALM LO
Allows the user to set the lower limit of the Diastolic
NIBP
alarm. (See
“NIBP Alarm Limits” on page 2-50.)
UNITS
Allows the user to select the unit of measurement for NIBP. The
selections are
mmHg
or
kPa
.
INTERVAL
Allows the user to select the interval at which NIBP measurement will
automatically initiate. It also allows the user to turn the automatic
measurement OFF. The selections are:
CONT
(continuous),
1min
,
2min
,
3min
,
4min
,
5min
,
10min
,
15min
,
30min
,
1HR
,
2HRS
,
4HRS
and
OFF
.
DISPLAY
TIMEOUT
Allows the user to select the interval at which the NIBP measurement
will time-out from the display. The selections are:
15min
,
30min
,
45min
, and
1HR
.
NIBP measurement will time-out from the display when the elapsed
time exceeds the chosen interval, except during an NIBP alarm
condition. The time-out will be suspended until the alarm condition is
resolved and will then resume. When the time-out interval has been
exceeded, the NIBP measurement is replaced by dashes.
RESTORE
DEFAULTS
Allows the user to restore the NIBP user default configuration.
FIGURE 2-39
Confirmation Dialog Box