Alarms
Operations
2 - 68
0070-10-0666-01
Trio™ Operating Instructions
ALM PRINT
Allows the user to enable or disable automatic printing during a
physiological alarm condition. Choose
ON
to enable printing
upon alarm. Choose
OFF
to disable printing upon alarm. See
“Recorder (Optional)” on page 2-82 for details.
SPO2 ALM HI
Allows the user to set the upper limit of the
SPO2
alarm. Choices
are 80 – 100% and OFF. Choose
OFF
to disable the alarm. If the
alarm is set to
OFF
, the alarm
OFF
symbol
will display within
the parameter tile. The alarm is activated when the measurement
equals or exceeds the high limit.
SPO2 ALM LO
Allows the user to set the lower limit of
SPO2
alarm.
PR ALM HI
Allows the user to set the upper limit of the
Pulse Rate
alarm.
PR ALM LO
Allows the user to set the lower limit of the
Pulse Rate
alarm.
SENSOR OFF
AUDIO
Controls the onset of the audio beep alarm for the “SpO2 Sensor
OFF” condition. The choices are
ON
and
OFF
.
If
ON
is selected, then the audio beep alarm will sound when an
“SpO2 Sensor OFF” condition occurs. If
OFF
is selected, then the
audio beep alarm will not sound when an “SpO2 Sensor OFF”
condition occurs.
SAT SECONDS
(Nellcor® only)
Controls pulse oximetry nuisance alarms (see section 2.4.3.2 for
details on SatSeconds™ features). The selections are
OFF
,
10
,
25
,
50
or
100
.
When SatSeconds is enabled, the SPO2 Low and High Alarm
functions are automatically disabled. As a safety precaution, if
three (3) SPO2 alarm violations occur within 60 seconds, a
priority 2 alarm will trigger even if the SatSeconds limit has not
been reached.