Trio™ Operating Instructions
0070-10-0666-01
6 - 9
Appendix
Patient Parameter Specifications
6.2.5
ECG Respiration Performance Requirements
6.2.6
NIBP Sub-System Performance Characteristics
The NIBP function is capable of providing systolic, diastolic, and mean blood pressure
measurements in Pediatric and Adult modes non-invasively, using a blood pressure cuff. It
operates over the pulse rate range of 40 to 240 bpm.
The NIBP function is in accordance with the requirements of EN 60601-2-30, EN 1060-1,
EN 1060-3 and ANSI/AAMI SP-10:1992.
Resolution:
1 bpm
Accuracy:
± 1 bpm or ± 1% whichever is greater.
Trigger Threshold Level:
200
μ
V (Lead II)
Trigger Indication:
There is an audible beep on every beat
captured
Sensing Leads:
Lead II
Source:
3-lead or 5-lead ECG Cable.
Range:
Adult: 6 to 120 breaths per minute
Pediatric: 6 to 150 breaths per minute
Accuracy:
± 2% or ± 2 breaths per minute, whichever
is greater
Excitation:
≤
300
μ
A RMS max.
Bandwidth:
0.2 Hz to 2 Hz (-3 dB) for all patient sizes
Baseline Impedance Range:
200
Ω
to 2500
Ω
at patient with 1 k
Ω
resistor in the ECG cable.
Linear Signal Range:
3
Ω
p-p minimum
Noise:
less than 0.05
Ω
at 500
Ω
patient
impedance, using a standard ECG cable
Min. Breath Height Detected:
Function of respiration scale. Waveform
needs to be greater than 0.3
Ω
in order for
breaths to be accurately detected.
Cardiovascular Artifact Rejection:
Detected by algorithm
Sweep speed:
6.25 mm/s, 12.5 mm/s, 25 mm/s