Trio™ Operating Instructions
0070-10-0666-01
2 - 47
Operations
Parameter Menus
2.4.4.3
NIBP Display Tile
If ECG and NIBP are selected in the
MODULE SETUP
menu, NIBP will be displayed in the
second parameter tile, beneath the ECG parameter tile. If ECG is not selected in the
MODULE SETUP
menu, NIBP will be displayed in the first parameter tile. The following
data is displayed:
FIGURE 2-37
NIBP Parameter Tile
The following status messages may be displayed during the measurement of NIBP:
MESSAGE
CAUSE
Cont measuring…
Displayed during Continuous measuring
Auto measuring…
Displayed during Automatic / Interval measuring
Please start
Displayed after selecting an
INTERVAL
in the
NIBP SETUP
menu
Measurement over
Displayed after an
NIBP
measurement has been manually stopped by
pressing the
NIBP
key during the measurement.
ET:5min
CUFF: 100
NIBP
mmHg
108
70
(84)
Cont measuring...
Interval:OFF
Systolic Value
Mean Value
Status Message
Elapsed Time
(since the last
measurement)
Diastolic Value
Unit of Measure
Current Cuff Pressure
Alarm OFF Symbol
Interval