Decontamination of the Monitor
User Maintenance
4 - 2
0070-10-0666-01
Trio™ Operating Instructions
4.2
Decontamination of the Monitor
WARNING: Perform the decontamination process with the unit powered
Decontamination of a unit that has come in contact with a biological material can be
performed using LpH SE Germicidal detergent. Apply a small amount of detergent to a
disposable wipe (paper based) and wipe down the outside of the unit. Discard the wipe
appropriately. After waiting 10 minutes, use a clean dry wipe to dry the unit.
CAUTION:
4.3
Care and Cleaning of the Monitor
The monitor housing may be cleaned with a mild soap and water solution or ammoniated
window cleaner. Apply cleaning solution to the cloth, not directly onto the monitor. DO NOT
apply large amounts of liquid. DO NOT use abrasive cleaning agents or organic solvents.
WARNING: Do not clean the monitor or sensors while it is on and/or
To prevent scratches on the screen, carefully brush dust and dirt particles with a soft sponge
moistened with cleaning solution; or a fine, soft-hair brush. Fingerprints and stains may be
removed by using a liquid lens cleaner and a soft cloth. DO NOT wipe a dry screen or use