Physical Specifications
Appendix
6 - 20
0070-10-0666-01
Trio™ Operating Instructions
6.3.7.1
Sealed Lead Acid Battery: P/N 0146-00-0043
6.3.7.2
Lithium Ion Battery: P/N 0146-00-0069
6.3.8
Data Storage
6.3.8.1
Monitor Configuration Data Storage
The monitor has the ability to store, in non-volatile memory, the user selectable configuration.
In addition to the factory default configuration for each patient size, there is 1 user
configurable configuration available for each patient size and the following information is
saved:
• Alarm Values
• Display Configurations
• NIBP Interval
• Parameter Settings, scaling
• Trace Speed
• Printer Settings
• Trend Configurations
6.3.8.2
Patient Data Storage
The current patient information and demographics are stored in non-volatile memory.
6.3.8.3
Patient Trend Data
The monitor is capable of storing a maximum of 24 hours of list and/or graphic trend values
for each active parameter. The minimum time interval for list trends is one minute. The
minimum interval for graphic trends is 1 second. When the monitor is powered OFF, the list
and graphic trend data is maintained for 2 hours. If the monitor remains OFF for more than 2
hours, the list and graphic trend data is deleted.
The minimum Battery Run Time is:
75 minutes from one fully charged new
battery at 25 °C for the following conditions,
which represent the most likely transportable
configuration: ECG, SpO
2
, and NIBP
running at the 15 minute interval.
The Battery Recharge Time is:
8 hours maximum
The minimum Battery Run Time is:
3.25 hours from one fully charged new
battery at 25 °C for the following conditions,
which represent the most likely transportable
configuration: ECG, SpO
2
, and NIBP
running at the 15 minute interval.
The Battery Recharge Time is:
6.5 hours maximum