Parameter Menus
Operations
2 - 44
0070-10-0666-01
Trio™ Operating Instructions
2.4.4
NIBP Monitoring
The
Trio
utilizes the oscillometric method of measuring Non-Invasive Blood Pressure (NIBP).
The measurement includes Systolic (SYS), Diastolic (DIA) and Mean Arterial Pressures (MAP).
Two (2) frequency modes of obtaining measurements are available:
MANUAL
and
INTERVAL
. Each mode will display the Systolic (SYS), Diastolic (DIA) and Mean Arterial
Pressure (MAP) values in the NIBP tile.
CAUTION:
Use only approved blood pressure cuffs and hoses with the
Trio.
CAUTION:
CAUTION:
Please consult a physician for interpretation of blood
pressure measurements.
CAUTION:
CAUTION:
CAUTION:
MANUAL MODE
Each time a measurement is desired, press the
NIBP
key on the
Trio
keypad to initiate a measurement. Press the
NIBP
key a second time
to stop a measurement already in progress.
INTERVAL MODE
Measurements are taken automatically at selected time intervals. The
selections are:
CONT
(continuous),
1min
,
2min
,
3min
,
4min
,
5min
,
10min
,
15min
,
30min
,
1HR
,
2HRS
,
4HRS
or
OFF
. The
NIBP
key on the
Trio
keypad must be pressed to initiate the first
measurement of the interval measurement cycle. If the patient is
discharged from the monitor while in interval mode, the interval
setting will revert to the User Default and the next measurement will be
delayed until the
NIBP
key is pressed. During the delay period, the
interval setting will remain displayed in the NIBP numeric tile and the
Elapsed Time (ET) will display as “0min”.