User Maintenance
Care and Cleaning of Accessories
4.4
Care and Cleaning of Accessories
4.4.1
SpO
2
Sensors
NOTE:
Refer to the individual instruction sheets that are packaged
with each sensor.
• Check the sensors and cables for signs of damage on a daily basis. Replace as required.
• The reusable sensors should be cleaned before and after each patient's use
• Wipe the patient contact area using a soft cloth with a mild soap and water solution or
isopropyl alcohol. Hydrogen peroxide can be used to remove dried blood from all
accessible surfaces
• The cable can be cleaned with a 3% hydrogen dioxide solution, isopropanol solution, or
other active reagent. However, the connector of the sensor is not to be subjected to such
a solution
• Allow the sensor to completely dry before using
CAUTION:
CAUTION:
Do not subject the sensor to autoclaving.
CAUTION:
CAUTION:
Do not use sensors or cables that are damaged or have
deteriorated.