Interval Real-Time Printout
The
Trio
can be set to print automatically at regular intervals, as designated by the user. To
activate the interval print feature:
1.
Select
INTERVAL
from the
PRINTER SETUP
menu.
2.
Use the Navigator™ Knob to choose the interval at which an automatic printout will
occur. The following time intervals are available:
10min
,
20min
,
30min
,
40min
,
50min
,
1HR
,
2HRS
,
3HRS
,
and
4HRS
.
3.
Select
OFF
to disable the automatic print function.
FIGURE 2-58
Example, Interval Real-Time Printout
Alarm Printout
When a parameter alarm is activated, and
ALM PRINT
is set to
ON
for the alarming
parameter, a printout is automatically generated for that parameter. In the
ALARM SETUP
menu, the user can designate the length of the waveform on the alarm printout as either
8
,
16
or
32
seconds.
To set the length of the alarm printout:
1.
Access the
SYSTEM MENU
using the
MENU
icon on the main display.
2.
Select
MONITOR SETUP
.
3.
Select
ALARM SETUP
.
4.
Scroll to
ALM PRINT TIME
and select a recording time length.
FIGURE 2-59
Example, Alarm Printout
Depending on the alarm print time selected, the print recording will display the waveform, 4,
8, or 16 seconds respectively, prior to and after the alarm event. For example, for an eight
(8) second alarm printout four (4) seconds of pre-alarm, and four (4) seconds of post alarm
data will appear on the printout. Alarm times of
16
or
32
seconds will print alarm data in a
similar manner. All numeric data displayed during the alarm event will also be included on