Trio™ Operating Instructions
0070-10-0666-01
2 - 11
Operations
System Menu
2.3.5.6
Trace Setup
TRACE SETUP
allows the user to choose the traces that will be displayed in the waveform
area and to choose the viewing mode. The viewing modes improve the viewability of the
display depending on how the
Trio
is positioned for use. There are two viewing modes that
can be selected from the
WAVE DISPLAY
pull-down menu.
•
MODE 1
– provides an enhanced view angle and a less smooth wave. This mode should
be used when the
Trio
is being viewed from an angle. In most instances, this mode will
provide optimal viewing of waveform data. (
MODE 1
is the factory default setting.)
•
MODE 2
– provides a basic view angle and a smooth wave. This mode should be used
when the
Trio
is wall-mounted and being viewed from below.
1.
Select
TRACE SETUP
from the
MONITOR SETUP
menu. The
TRACE SETUP
menu
FIGURE 2-10
Trace Setup Menu
2.
Use the Navigator™ Knob to select each parameter waveform (a check mark in the box
next to a parameter indicates that it has been selected). To deselect a parameter
waveform, press the knob again.
3.
Choose the desired viewing mode from the
WAVE DISPLAY
pull-down menu.
4.
Select
PREVIOUS MENU
to return to the previous menu. Select
NORMAL SCREEN
(from the menu or the Front Panel Keypad) to exit the menu and return to the normal
screen.
NOTE:
If the optional IBP is not installed, it will not be provided as
a choice in the TRACE SETUP menu.