Trio™ Operating Instructions
0070-10-0666-01
2 - 59
Operations
Parameter Menus
IBP Setup Menu Selections
ALM
Allows the user to turn the
IBP
alarm
ON
or
OFF
. Choose
ON
to
enable the alarm; choose
OFF
to disable the alarm. If the alarm is set
to
OFF
, the alarm
OFF
symbol
will display to the right of
IBP
on
the screen. The
IBP
alarm is activated when the blood pressure value
exceeds set high or low
IBP
values.
ALM PRIORITY
Enables priority selection for the
IBP
alarm. Choices are
1
,
2
and
3
.
Priority 1 alarms are considered the most serious.
ALM PRINT
Enables or disables automatic printing during an
IBP
alarm condition.
Choose
ON
to enable report printing upon
IBP
alarm. Choose
OFF
to disable printing upon
IBP
alarm.
SYS ALM HI
Allows the user to set the upper limit of the Systolic
IBP
alarm.
(See “IBP Alarm Limits” on page 2-61.)
SYS ALM LO
Allows the user to set the lower limit of the Systolic
IBP
alarm.
(See “IBP Alarm Limits” on page 2-61.)
MEAN ALM HI
Allows the user to set the upper limit of the Mean Arterial Pressure
IBP
alarm. (See “IBP Alarm Limits” on page 2-61.)
MEAN ALM LO
Allows the user to set the lower limit of the Mean Arterial Pressure
IBP
alarm. (See “IBP Alarm Limits” on page 2-61.)
DIA ALM HI
Allows the user to set the upper limit of the Diastolic
IBP
alarm.
(See “IBP Alarm Limits” on page 2-61.)
DIA ALM LO
Allows the user to set the lower limit of the Diastolic
IBP
alarm.
(See “IBP Alarm Limits” on page 2-61.)
SWEEP
Adjusts the speed of the IBP waveform on the display. The selections
are
12.5
and
25.0
mm/sec.
UNITS
Allows the user to select the units of measurement for IBP. The
selections are
mmHg
or
kPa
.