Trio™ Operating Instructions
0070-10-0666-01
2 - 25
Operations
Parameter Menus
ECG Setup Menu Selections
ALM
NOTE:
In the French configuration, the HR
(Heart Rate) alarm is always ON. It
cannot be disabled by turning it OFF.
Allows the user to turn the HR (Heart Rate) alarm
ON
or
OFF
. Choose
ON
to enable the alarm; choose
OFF
to disable the alarm. If the
alarm is set to
OFF
, the alarm
OFF
symbol
will display to the right
of
ECG
on the screen. The HR alarm is activated when the heart rate
is equal to or exceeds set high or low HR values.
ALM PRIORITY
Allows the user to select the priority of the ECG alarm. Choices are
1
,
2
and
3
. Priority 1 alarms are considered the most serious.
ALM PRINT
Enables or disables automatic printing during a
HR
alarm condition.
Choose
ON
to enable printing upon HR alarm. Choose
OFF
to
disable printing upon HR alarm.
HR ALM HI
Allows the user to set the upper limit of the
HR
Rate Alarm Limits” on page 2-27.)
HR ALM LO
Allows the user to set the lower limit of the
HR
Rate Alarm Limits” on page 2-27.)
SOURCE
Determines the source of heart rate. An audible tone is generated
when a heart beat is detected from the selected source. The selections
are
ECG
,
SpO2
and
AUTO
.
NOTE:
When heart rate is being measured, a
heart rate source label
(“SOURCE:XXXX”) is displayed in the
ECG parameter tile. This label indicates
from where the heart rate parameter
is derived and is displayed in the same
color as the ECG waveform.
ECG:
When ECG is selected as the heart rate source, the label
HR
(heart rate) is displayed in the
ECG
parameter tile. This label and its
associated numeric value are displayed in the same color as the ECG
waveform. The audible tone sounds with each R wave.
SpO2:
When SpO
2
is selected as the heart rate source, the label
PR
(pulse rate) is displayed in the
ECG
and
SpO2
parameter tiles. This
label and its associated numeric value are displayed in the same color
as the SpO
2
numeric value. The audible tone sounds at the peak of
each pulse wave.