Trio™ Operating Instructions
0070-10-0666-01
3 - 3
Defaults
Default Configurations
Print Grid
ON
ANALOG SETUP
Analog Output Enable
OFF
Analog Wave
ECG
PARAMETER COLORS
ECG - green
PLETH - blue
RESP - yellow
NIBP/TEMP - white
MODULE SETUP in
MONITOR SETUP
Will be the same as
MODULE SETUP submenu
of the FACTORY menu.
TRACE SETUP
ECG, SpO2, RESP, IBP (if
IBP is set to display from
the MODULE SETUP
submenu of the FACTORY
menu)
Wave Display
Mode 1
ECG SETUP
ECG HR alarm
Standard configuration
French configuration
OFF
ON
ECG HR Alarm Print
OFF
ECG HR Alarm Priority
1
HR Source
ECG
HR Cascade
ON
ECG Lead Type
5 Lead
ECG Sweep Speed
25 mm/s
ECG Beep Volume
Medium
ECG Pacer
OFF
ECG Filter Mode
Extended
ECG Ch1 Lead Mode
II
ECG Ch1 Gain
x2
ECG HR Alarm Limit Hi
150
175
ECG HR Alarm Limit Lo
45
70
RESP SETUP
Resp Rate Alarm
OFF
Resp Rate Alarm Print
OFF
Resp Rate Alarm Priority
2
Resp Sweep Speed
12.5 mm/s
Resp Wave Scale
1
Resp Rate Alarm Limit Hi
30
Resp Rate Alarm Limit Lo
6
SPO2 SETUP
SpO2 Alarm Print
OFF
SpO2 Alarm Priority
2
CONFIGURATION ITEM
FACTORY
DEFAULTS
ADULT
(Patient size)
FACTORY
DEFAULTS
PEDIATRIC
(Patient size)
FACTORY
DEFAULTS
COMMON